Product Specifications
A-13
Accuracy
±3 bpm (measured without motion)
±5 bpm (measured with motion)
Refreshing rate 1s
Low Perfusion
Conditions
Pulse amplitude: >0.02%
Light penetration: >5%
SpO
2
accuracy ±2%
PR accuracy ±3 bpm
Nellcor SpO
2
Module
Parameter Specifications
Sensor Range Precision*
MAX-A, MAX-AL, MAX-N
MAX-P, MAX-I, MAX-FAST
70 to 100%
0% to 69%
±2%
Not specified.
OxiCliq A, OxiCliq N
OxiCliq P, OxiCliq I
70 to 100%
0% to 69%
±2.5%
Not specified.
D-YS, DS-100A, OXI-A/N,
OXI-P/I
70 to 100%
0% to 69%
±3%
Not specified.
SpO
2
measurement range and
accuracy
MAX-R, D-YSE, D-YSPD
70 to 100%
0% to 69%
±3.5%
Not specified.
20 to 250 bpm: ±3 bpm
PR measurement range and
accuracy
251 to 300 bpm: Not specified.
Refreshing rate 1s
*: When the SpO
2
sensor is applied for neonatal patients as indicated, the specified accuracy range
is increased by ±1%, to compensate for the theoretical effect on oximeter measurements of fetal
hemoglobin in neonatal blood.