EasyManuals Logo

Mindray BeneView T8 Operation Manual

Mindray BeneView T8
290 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #244 background imageLoading...
Page #244 background image
Product Specifications
A-12
A.5.3 SpO
2
MPM SpO
2
Module
Standards Meet standards of ISO9919 and ASTM1415
SpO
2
Measurement range 0 to 100%
Resolution 1%
Accuracy
70 to 100%: ±2% (measured without motion in adult/pediatric
mode)
70 to 100%: ±3% (measured with motion in neonate mode)
70 to 100%: ±3% (measured with motion)
0% to 69%: Not specified.
Refreshing rate 1s
PR
Measurement range 20 to 254 bpm
Resolution 1bpm
Accuracy
±3 bpm (measured without motion)
±5 bpm (measured with motion)
Refreshing rate 1s
Masimo SpO
2
Module
SpO
2
Measurement range 1 to 100%
Resolution 1%
Accuracy
70 to 100%: ±2% (measured without motion in adult/pediatric
mode)
70 to 100%: ±3% (measured with motion in neonate mode)
70 to 100%: ±3% (measured with motion)
0% to 69%: Not specified.
Refreshing rate 1s
PR
Measurement range 25 to 240 bpm
Resolution 1 bpm

Table of Contents