Product Specifications
A-12
A.5.3 SpO
2
MPM SpO
2
Module
Standards Meet standards of ISO9919 and ASTM1415
SpO
2
Measurement range 0 to 100%
Resolution 1%
Accuracy
70 to 100%: ±2% (measured without motion in adult/pediatric
mode)
70 to 100%: ±3% (measured with motion in neonate mode)
70 to 100%: ±3% (measured with motion)
0% to 69%: Not specified.
Refreshing rate 1s
PR
Measurement range 20 to 254 bpm
Resolution 1bpm
Accuracy
±3 bpm (measured without motion)
±5 bpm (measured with motion)
Refreshing rate 1s
Masimo SpO
2
Module
SpO
2
Measurement range 1 to 100%
Resolution 1%
Accuracy
70 to 100%: ±2% (measured without motion in adult/pediatric
mode)
70 to 100%: ±3% (measured with motion in neonate mode)
70 to 100%: ±3% (measured with motion)
0% to 69%: Not specified.
Refreshing rate 1s
PR
Measurement range 25 to 240 bpm
Resolution 1 bpm