Product Specifications
A-19
A.5.9 AG
Standards Meet standard of ISO 21647
Technique Infrared absorption
Warm-up time
Warming up status:
Ready-to-measure status:
45 s
10 min
Sample flow rate
Adult, pediatric:
Neonate:
Accuracy:
120, 150, 200 ml/min
70, 90, 120 ml/min
±10 ml/min or ±10%, whichever is greater
Measurement range
CO
2
:
O
2
:
N
2
O:
Des:
Sev:
Enf:
Iso:
Hal:
awRR:
0 to 30%
0 to 100%
0 to 100%
0 to 30%
0 to 30%
0 to 30%
0 to 30%
0 to 30%
2 to 100 rpm
Resolution
CO
2
:
awRR:
1 mmHg
1 rpm
Accuracy
(warming-up status)
CO
2
:
N
2
O:
Other anesthetic gases:
±0.3%
ABS
±(8%
REF
+2%
ABS
)
8%
REF
Gases Range (%
REL
) Accuracy (%
ABS
)
0 to 1 ±0.1
1 to 5 ±0.2
5 to 7 ±0.3
7 to 10 ±0.5
CO
2
>10 Not specified
0 to 20 ±2
N
2
O
20 to 100 ±3
0 to 25 ±1
25 to 80 ±2
O
2
80 to 100 ±3
0 to 1 ±0.15
1 to 5 ±0.2
Accuracy
(ready-to-measure
status)
Des
5 to 10 ±0.4