Product Specifications
A-21
A.5.10 ICG
Technique Thoracic electrical bioimpedance (TEB)
Measurement range
SV:
HR:
C.O.:
5 to 250 ml/beat
44 to 185 bpm
1.4 to 15 L/min
Accuracy
SV:
HR:
C.O.:
Not specified.
±2 bpm
Not specified.
A.5.11 BIS
Standards Meet standard of IEC 60601-2-26
Technique Bispectral index
Measured parameters
EEG
BIS: 0 to 100
Calculated parameters
SQI: 0 to 100%
EMG, SR, DSA, TP
Impedance range 0 to 999 kΩ
Sweep speed 12.5, 25 or 50 mm/s
Input impedance >5 MΩ
Noise (RTI) <0.3 µV (0.25 to 50 Hz)
Input signal range ±1 mV
EEG bandwidth 0.25 to 110 Hz
Patient leakage current <10 µA