Product Specifications
A-22
A.5.12 RM
Technique Flow sensor
Frequency response ≥30Hz
Dead space <11ml
Flow
Measurement range
Adult/pediatric:
Neonate:
± (2 to 120) L/min
± (0.5 to 30) L/min
Accuracy 25 ml/s or ±10% of the reading, whichever is greater
Resolution 0.1 L/min (minimum)
Paw
Measurement range -20 to 120 cmH
2
O
Accuracy ±3% of the reading
Resolution 0.1 cmH
2
O
MVe/MVi
Measurement range
Adult, Pediatric:
Neonatal:
2 to 60 L/min
0.5 to 15 L/min
Accuracy ±5%×reading
TVe/TVi
Measurement range
Adult, Pediatric:
Neonatal:
100 to 1500 ml
20 to 500 ml
Resolution 1 ml
Accuracy ±15 ml or ±10%×reading, whichever is greater.
RR (RM)
Measurement range 4 to 120 rpm
Accuracy
4 to 99 rpm
100 to 120 rpm
±1 rpm
±2 rpm
Calculated Parameters
Measurement range Measurement accuracy
I:E 12:1 to 1:12
FEV1.0% 0 to 100%
Pmean 0 to 120 cmH
2
O
TV 20 to 1500 ml
MV 2 to 60 L
PEEP 0 to 120 cmH
2
O
<±5% of the reading