DICOM 12-9
z Select an item in the service list, and click [Delete] to delete the service.
z Select an item in the service list, and click [Default] to set the server to be the
default service.
2. Select an item in the service list, and click [Verify] to verify the connection.
3. Click [Exit] to confirm the setting and exit the page.
DICOM service setting for Worklist is described as follows:
Name Description
Device
After you set the server(s) in DICOM Server Setting screen, the
name(s) will appear in the drop-down list, select the name of the
Worklist server.
Service Name Default is xxx-Worklist, and it can be modified.
AE Title
Application Entity title, Here, it should be consistent with that of the
Worklist server.
Port
DICOM communication port, 104 by default. Here, the port should be
consistent with that of the Worklist server port.
Maximum Retries Reserved feature
Interval Time(Sec) Reserved feature.
Timeout(Sec)
Refers to time after which the system will stop trying to establish a
connection to the service. Value: 560s, in increments of 5s, and 15
by default.