Probes and Biopsy 13-15
For details about additional infection control procedures or disinfection
information, contact
Mindray North America Customer Service [ Tel: (1-800) 288 2121, (1-201) 995
8000 ].
13.1.6 Storage and Transportation
When all examinations for the day have been completed, confirm that the probe is in good
condition. After disinfecting the probe, confirm that the probe is in good condition and
store it in a suitable place so that the next examination can be conducted smoothly.
1. To prevent the probe from being damaged, DO NOT store it in locations where it may
be exposed to:
z Direct sunlight or X-rays
z Sudden changes in temperature
z Dust
z Excessive vibration
z Heat generators
2. Store and transport the probe under the following ambient conditions:
z Ambient temperature: -20°C ~ 55°C
z Relative humidity: 30% ~ 95% (no condensation)
z Atmospheric pressure: 700 hPa ~ 1060 hPa
Store and transport the 4CD4 probe under the following ambient conditions:
z Ambient temperature: -10°C ~ 60°C
z Relative humidity: 30% ~ 95% (no condensation)
z Atmospheric pressure:700 hPa ~ 1060 hPa
3. When the probe is sent to MINDRAY Customer Service Department or sales
representative for repair, be sure to disinfect it and keep it in the carrying case to
prevent infection.
4. Sterilize the carrying case as necessary.
13.2 Biopsy Guide
WARNING:
1. The person performing biopsy procedures must
understand diagnostic ultrasound thoroughly and have
been trained adequately, otherwise, side effects may be
caused to the patient.
2. In situations listed below, the biopsy needle may fail to
penetrate the target. The incorrect biopsy may cause
various side effects in the patient.
Use a needle-guided bracket other than that provided.
Mount the needle-guided bracket incorrectly.
Use a biopsy needle that is unsuitable for the type of
biopsy being performed.
Use a biopsy needle or guiding block that is unsuitable
for the needle guide.