13-32 Probes and Biopsy
CAUTION:
Ensure that all guide parts are seated properly prior to
performing a biopsy.
13.2.5 Biopsy Menu
Move the cursor onto the image menu; navigate the cursor to [Other] item and press
<Set> or the multifunctional knob. Click [Biopsy] to enter the biopsy menu.
Select biopsy bracket angle
If the needle-guided bracket supports more than one biopsy angle, you can select the
angle from the drop-down list (the last two numerals).
Select Guide Line
If a bracket has more than one guide lines, then you can see the [Guide Line] item,
and click to display the guide lines, whether one line or all.
Display or hide the guide line
To display the guide line, set [Display] as “On”; to hide the guide line, set [Display] as
“Off”.
Or, press <Biopsy> on the control panel to display or hide the guide line.
Select guide line dot size
Move the cursor onto [Dot Size], press <Set> to select the dot size among Small,
Medium and Big.
Tips:
z The guide line is a dot line which consists of two kinds of dots, the distance
between two dots is depth dependent. Move the cursor onto the big dot, a
numeral, which represents the biopsy depth, is displayed.
z The biopsy guidezone adjusts along with image adjustments, such as image
inversion/rotations, zoom and depth changes.
z When the imaging depth and scan range are changed, the guide line will be
adjusted.
Select through B mode image menu:
z [Biopsy]: to display or hide the guide line.
z [Biopsy Kit]: to select bracket angle.
Preset the two items through: [Setup][Image Preset][Menu Preset].