Safety Precautions 1-9
8.
Please use the disinfection or sterilization solution that
recommended in this operator’s manual; otherwise
Mindray will not be liable for damage caused by other
solutions. If you have any questions, please contact
Mindray Customer Service Department or sales
representative.
9.
The probe sheath contains natural rubber that can cause
allergic reactions in some individuals.
10.
Do not use pre-lubricated condoms as a sheath.
Lubricant may not be compatible with the transducer
material and damage may result.
11.
Transducer damage may be caused by inappropriate gel,
detergent or cleanser:
Do not soak or saturate transducers with solutions
containing alcohol, bleach, ammonium chloride
compounds, acetone or formaldehyde.
Avoid contact with solutions or coupling gels containing
mineral oil or lanolin.
NOTE:
1. Read the following precautions to prevent the probe from malfunction:
z Clean and disinfect the probe before and after each examination.
z After the examination, wipe off the ultrasound gel thoroughly.
Otherwise, the ultrasound gel may solidify and the image quality
would be degraded.
2. Ambient conditions:
To prevent the probe from being damaged, do not use it where it will be
exposed to:
z Direct sunlight or X-rays
z Sudden changes in temperature
z Dust
z Excessive vibration
z Heat generators
Use the probes under the following ambient conditions:
z ambient temperature: 0℃~ 40℃
z relative humidity:30% ~ 85% (no condensation)
z atmospheric pressure:700 hPa ~ 1060 hPa.
Use the 4CD4 probe under the following ambient conditions:
z ambient temperature: 10°C ~ 40°C
z relative humidity: 30% ~ 85% (no condensation)
z atmospheric pressure: 700 hPa ~ 1060 hPa
3. Repeated disinfection will eventually damage the probe, please check the
probe's performance periodically.