5.6 Command Explanation
5-45 ZM2371/ZM2372
BUF1, BUF2, BUF3
<measured data 1>,<measured data 2>,<measured data 3>,…,<measured
data n>
“n” is the measurement point number specified by “:DATA:POINts”
command.
The read data is deleted from the measured data buffer.
The parts not recorded are all measured data zero (for example,
+0,+0.00000E+00,+0).
The content of <measured data m> is as follows (m: 1 to n).
<measurement status>, <measured data>
[ , {comparator’s bin sorting result>|<limit comparison result>} ]
<measurement status> {numeric value, format NR1, range see below}
0 No error
1 Either measurement error ERR, ALC error ALC Err, or correction
error CORR Err
2 Contact failure
3 Other any errors
<measured value>
For BUF1 and BUF2, either <primary parameter measured value> or
<secondary parameter measured value> set by :DATA:FEED
command.
For BUF3, <primary parameter measured value> and <secondary
parameter measured value>.
For data transfer format, :FORMat[:DATA] command
<comparator’s bin sorting result> {numeric value, format NR1, range see
below}
For BUF1 and BUF2, the bin sorting result is output whichever the
comparator function (bin sorting) is enabled or disabled.
For BUF3, the bin sorting result is output only when the comparator
function (bin sorting) is enabled.
When bin extended function When bin extended function
is disabled is enabled
0 OUT_OF_BINS 0 OUT_OF_BINS
1 BIN1 1 BIN1
2 BIN2 2 BIN2
• • • •
9 BIN9 9 BIN9
10 AUX_BIN(S-NG) 10 BIN10
11 Sorting failed 11 BIN11
• •
14 BIN14
15 AUX_BIN(S-NG)
16 Sorting failed
AUX_BIN (S-NG): Output only when S-NG is used as auxiliary bin.
:CALCulate:COMParator:AUXBin command
Sorting failed: Any error occurred or comparator function is disabled.
When the bin extended function is enabled, the range exceeding 10
has different meaning.
:CALCulate:COMParator:EXTension[:STATe] command