5.6 Command Explanation
5-26 ZM2371/ZM2372
*RCL <memory number>
Description: Recalls the settings from the specified settings memory.
Parameters: <memory number> {numeric value, range 0 - 31}
Out of range value causes an error.
Setting example: *RCL 5
Recalls the settings from the settings memory number 5.
Remarks: To recall correction values, use the :SYSTem:RCL command.
*RST
Description: Resets to initial setting state.
For details • • • “Table 3-1 Setting items and initial values”
Setting example: *RST
Remarks: Beware of the following points:
• OPEN, SHORT, and LOAD correction values are initialized. When
using previous correction values, save them in advance, and recall
them after *RST.
• • • :SYSTem:SAVE, :SYSTem:RCL commands
• The initial value of trigger delay time is not zero.
*SAV <memory number>
Description: Saves the currently used setting to the specified setting memory.
Parameters: <memory number> {numeric value, range 0 - 31}
Out of range value causes an error.
Setting example: *SAV 5
Saves current settings in the settings memory number 5.
Remarks: To save correction values, use the :SYSTem:SAVE command.
*SRE <SRQ mask>
*SRE?
Description: Sets/queries the service request enable register.
Parameters: <SRQ mask> {numeric value, range 0 - 255}
Out of range value causes an error.
For details • • • “5.7.1 Status System Overview”
Setting example: *SRE 128
Response: {numeric value, format NR1, range 0 - 255}
Query example: *SRE?
Response example: +128
The content of service request enable register is 128.
*STB?
Description: Queries the content of status byte register.
Response: <register contents> {numeric value, format NR1, range 0 - 255}
For details • • • “5.7.2 Status Byte”
Query example: *STB?
Response example: +128
The content of status byte register is 128.