5.6 Command Explanation
5-48 ZM2371/ZM2372
:DATA:POINts {BUF1|BUF2|BUF3},<buffer size>
:DATA:POINts? {BUF1|BUF2|BUF3}
Description: Sets/queries virtual size (number of measurement points) of the
measured data buffer.
If set, the specified measured data buffer is cleared.
Parameters: BUF1 Sets/queries the measured data buffer 1.
BUF2 Sets/queries the measured data buffer 2.
BUF3 Sets/queries the measured data buffer 3.
<buffer size> Expressed with the number of measurement points.
MAX / MIN can be used.
BUF1, BUF2 {numeric value, range 1 - 200}
BUF3 {numeric value, range 1 - 1000}
Setting example: :DATA:POIN BUF1,100
Sets virtual size of measured data buffer 1 to 100.
Response: BUF1, BUF2 {numeric value, format NR1}
BUF3 {numeric value, format NR1}
Query example: :DATA:POIN? BUF1
Response example: +100
Virtual size of measured data buffer 1 is 100.
Remarks: The measured data buffer is a ring buffer. After it becomes physically
full, the data in the lead position is overwritten. When the measured
data are recorded by the amount of virtual size, the status that
indicates “full” is set. • • • 5.7.4 Operation Status
For :DATA[:DATA]? {BUF1|BUF2|BUF3}, the data by the amount of
virtual size are transmitted starting from the lead position, whichever
the buffer is full or not.
:DISPlay[:WINDow][:STATe] <display permission>
Description: This command is accepted but it does nothing.
Parameters: <display permission > {ON|OFF|1|0}
Setting example: :DISP OFF
Nothing is done. The measured result is always displayed.
:DISPlay[:WINDow]:TEXT1:DIGit <display digits>
Description: This command is accepted but it does nothing.
Parameters: < display digits > {numeric value, range 3 - 6}
Out of range causes an error.
Setting example: :DISP:TEXT1:DIG 4
Nothing is done. The measured value is always displayed with the
maximum of digits.
:DISPlay[:WINDow]:TEXT1:PAGE <displayed data>
Description: This command is accepted but it does nothing.
Parameters: < displayed data>{numeric value, range 1 or 2}
Out of range causes an error.
Setting example: :DISP:TEXT1:PAGE 2
Nothing is done.
The comparator result and limit comparison result are displayed
simultaneously with the measured value.