EasyManua.ls Logo

NF ZM2371 - Page 176

NF ZM2371
255 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5.6 Command Explanation
5-49 ZM2371/ZM2372
:DISPlay[:WINDow]:TEXT2:PAGE <display item>
Description: Sets auxiliary display items. Converted into :DISP:TEXT3 command
and executed.
Parameters: <display item> {numeric value, range 1 - 8}
Out of range causes an error.
2 Measurement condition status
4 Lower limit value and upper limit value of BIN1 (primary
parameter)
5 Lower limit value and upper limit value of secondary
parameter
7 Voltage & current monitor
1, 3, 6, 8 Content of auxiliary display is not changed but previous
display item is retained.
Setting example: :DISP:TEXT2:PAGE 4
Sets so as to display the lower limit value and upper limit value of BIN1
(primary parameter).
:DISPlay[:WINDow]:TEXT3[:PAGE] {STATus|BIN1|..|BIN14|SLIMit|REFerence|
MONitor}
:DISPlay[:WINDow]:TEXT3[:PAGE]?
Description: Sets/queries auxiliary display item.
Parameters: STATus Measurement condition status
BIN1..BIN14 Lower limit value and upper limit value of the specified
bin of primary parameter
SLIMit Lower limit value and upper limit value of secondary
parameter (Secondary LIMit)
REFerence Reference values of primary parameter and secondary
parameter (for deviation)
MONitor Current and voltage monitored values
Setting example: :DISP:TEXT3 MON
Sets auxiliary display item to the voltage & current monitor.
Response: {STAT|BIN1|..|BIN14|SLIM|REF|MON}
Query example: :DISP:TEXT3?
Response example: MON
Auxiliary display item is the voltage & current monitor.
Remarks: To display the lower limit value and upper limit value of primary
parameter used for the limit comparison function, specify the BIN1.
Do not specify a bin by mistake which is not used for the limit
comparison.

Table of Contents

Related product manuals