3.5 Basic Operations
3-31 ZM2371/ZM2372
3.5.6 Setting of Basic Measurement Conditions
3.5.6.1 Measurement Frequency
The measurement frequency is displayed at the upper right of measurement screen.
1.0000k Measurement frequency Hz
1.00 V_ Measurement signal level
_
FREQ
_
Press the
_
FREQ
_
key to display the measurement frequency setting menu.
Frequency: 1.0000kHz Current setting
0)120 1)1k 2)10k 3)100k 4)Entry Options
120 120Hz
1k 1kHz
10k 10kHz
100k 100kHz
Entry Displays the measurement frequency input menu to input a numeric
value.
If a setting is made or the EXIT operation is performed, the measurement screen comes
back.
Entry: The measurement frequency input menu is as follows.
Frequency: 1.0000kHz Current setting
0.001Hz to 100.00kHz Range of settable values
The setting resolution is 5 digits (1mHz when < 10Hz).
If a setting is made or the EXIT operation is performed, one-previous menu comes back.
_
FREQ
_