5.6 Command Explanation
5-59 ZM2371/ZM2372
[:SENSe]:CORRection:DATA {STANdard1|STANdard2|STANdard3} ,
<correction primary parameter>, <correction secondary parameter>
[:SENSe]:CORRection:DATA? {STANdard1|STANdard2|STANdard3}
Description: Sets/queries spot OPEN correction value, spot SHORT correction value,
and LOAD correction value at current frequency.
Set the format of correction value with
[SENSe]:CORRection:CKIT:STANdard{1|2|3}:FORMat command.
Parameters: STANdard1 Sets/queries spot OPEN correction value.
STANdard2 Sets/queries spot SHORT correction value
STANdard3 Sets/queries LOAD correction value.
<correction primary parameter> {numeric value, range 0,±(1E-16 -
9.99999E+11) }
<correction secondary parameter> {numeric value, range 0,±(1E-16 -
9.99999E+11) }
Actually set value follows the format of correction value and it is
rounded to the range same as set by the panel operation. The unit
follows the format of correction value.
Setting example: :CORR:DATA STAN2, 15.3E-03, 2.35E-03
Sets spot SHORT correction value to Rs = 15.3 m, X = 2.35 m.
(when the format of correction value is RX)
Response: <correction primary parameter>, <correction secondary parameter>
{numeric value, format NR3}
Query example: :CORR:DATA? STAN2
Response example: +1.53000E-02,+2.3500E-03
Spot SHORT correction value is Rs = 1.53000E-02 (15.3 m),
X = 2.3500E-03 (2.35 m). (when the format of correction value is RX)