4 Safety Instructions
Instruction Manual Keratograph 5M (G/77000/XXXX/EN – Rev05) 13 / 106
4.2 Safety Instructions for Use
Attention
Personal injury or property damage due to improper operation
Observe the following safety instructions.
Attention
Personal injury or property damage due to device modification
No modifications may be made to this device without the permis-
sion of the manufacturer. Only the OCULUS service are allowed
to modify the device or the associated lifting table
to install software and software updates.
Any serious-incident that has occurred in relation to the device should
be reported to the manufacturer (vigilance@oculus.de) and the compe-
tent authority of the region in which the user and/or patient is estab-
lished.
4.3 Instructions for Operating Personnel
Make certain that the Keratograph 5M is used exclusively in clinics
and by eye specialists and opticians (trained staff etc.).
For this reason the Keratograph 5M may only be operated by
personnel instructed to do so, who, with appropriate training,
knowledge and practical experience, are able to ensure proper
handling of the device.
4.4 Transport and Storage Instructions
Refer to the notes in Chap. 23, page 88.
4.5 Instructions for Setup and Connection
Only OCULUS or an authorized dealer is allowed to set up and to
connect the Keratograph 5M.
Do not use or store the Keratograph 5M in rooms that are humid,
Chap. 23, page 88.
Keep the Keratograph 5M away from water that may drip, splash or
spray on it, and make sure that no liquids can get into the
Keratograph 5M. Do not place any containers holding liquids in the
vicinity of the Keratograph 5M.
Only operate the Keratograph 5M in rooms used for medical pur-
poses after they have been set up according to the VDE Regula-
tions 0100-710.
Do not operate the devices included in the delivery in areas where
explosions may occur, or in proximity to flammable anesthetics or
volatile substances such as alcohol, benzine or similar products.
Set up the Keratograph 5M so that the power plug is easy to ac-
cess. That way, you can easily disconnect it from the power supply
for any repairs or maintenance work.