4 Safety Instructions
14 / 106 Instruction Manual Keratograph 5M (G/77000/XXXX/EN – Rev05)
Do not use excessive force when connecting the electrical plug.
If you are unable to make a plug connection, check whether the
plug fits the socket.
If you detect damage to the connection, you should let authorized
dealer repair the damage.
Only use a Keratograph 5M which is mounted at the lifting table
properly.
4.6 Patient environment information
Patient environment is the area where patients can come into contact
with any part of a medical electrical equipment (ME equipment) or with
another person being in contact with the ME equipment.
Attention
In the patient environment, use only devices that conform to IEC 60601-
1. If a multiple power socket is to be used, or if a device is to be used
that does not meet the IEC 60601-1 standard, use an isolating trans-
former.
Fig. 4-1: Patient environment