EasyManua.ls Logo

OCULUS Keratograph 5M - Page 84

OCULUS Keratograph 5M
104 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
21 Cleaning, disinfection and maintenance
86 / 106 Instruction Manual Keratograph 5M (G/77000/XXXX/EN – Rev05)
Align the sphere holder vertically so that the reference sphere is at
the same height as the black marking (ring) on the left vertical
brace of the head-chin rest.
Ensure that the reference sphere is still aligned parallel to the
device.
In the [Settings] menu, select the [Reference Measure] menu item.
Perform a measuring operation with the reference sphere
(Chap. 10.6, page 34).
Confirm the question “Calibration ok” with [OK].
Note
If the error message “Reference sphere not completely measured!” ap-
pears, the sphere must be carefully cleaned and the measuring opera-
tion repeated.
The system is now ready for operation. The reference data is stored di-
rectly on the device, so that the measuring head is not dependent on a
particular computer or laptop.
Fig. 21-5: Align the height of the reference sphere
1 Marking for height alignment

Table of Contents

Related product manuals