134
Chapter 8 Storage and maintenance
4 Turn the CO
2
cartridge in the direction of the arrow to remove it
from the pressure regulator.
5 See "When using a CO
2
cartridge" (page 37) for information about
attaching a new CO
2
cartridge.
CAUTION
Replacement of the CO
2
cartridge should only be done after confirming that the CO
2
cartridge is completely empty. Turn the cartridge dial in the OPEN direction until it stops
turning, loosen it completely and then check that the pressure indicator gauge is at zero.
Do not remove the CO
2
cartridge if there is any gas in it. The high pressure gas inside it
would be expelled.