8
The distal end of the insertion tube may become hot due to the illumination light or the
internal heating caused by electrical parts. Do not forget to turn off the light source before
attaching or detaching the optical adapter. Touching the distal end immediately after having
turned off the light source may burn your skin.
Do not allow the light emitted from the insertion tube's distal end to enter your eyes directly.
Otherwise, an eye injury may result.
Should any abnormality be detected during angulation operation, do not continue the
angulation operation by force. Otherwise, the insertion tube or examination target may be
damaged.
Do not subject the LCD surface to impact or strong pressure. Doing so could shatter the LCD
screen and cause personal injury.
Do not allow a metallic object or a fluid such as water enter the battery compartment, DC IN
connector or RECORDING CARD slot. Should any foreign object enter the instrument,
remove the battery and the AC adapter from the instrument and immediately contact
Olympus.
Do not touch the connectors directly by hand. Otherwise, a malfunction or electric shock may
result.
Do not allow metallic or foreign objects to enter the device through connectors or any other
openings. Otherwise, a malfunction or electric shock may result.
In case an irregularity such as smoke, abnormal odor or abnormal noise is detected,
immediately stop using the equipment even when it looks operable, and do not turn it ON
again.
On connecting this product, some GFCI safety plug may work its safety circuit up to its
threshold. Also, a GFCI safety plug does not prevent electrical shocks if the exterior of the
insertion tube contacts a live current.
Do not force an air connector that does not fit onto the air supply connector.
Do not supply air that is over the rating for the air supply connector.
Supply only air or CO
2
to the air supply connector, do not supply any liquid or flammable gas.
Otherwise, an explosion or fire may result.
When using a CO
2
cartridge, be sure to work only within the operating environment of the
CO
2
cartridge.
Always remove the CO
2
cartridge before transporting (air or land) the instrument.
Handle the CO
2
cartridge according to your local regulations.
CAUTION
The image quality may be deteriorated in high temperature.
Whenever the red temperature indicator and a temperature warning message appears along
with an audible warning alarm, withdraw it from the observation object immediately.
Otherwise the insertion tube and optical adapter may be damaged, or any other defects such
as malfunction or illumination dim may occur.
Do not insert the insertion tube in a machine that is running, as this may result in damage to
the insertion tube and/or machine. Also do not run the machine while the insertion tube is
inserted in it.
The minimum bending radius of the insertion tube is 60 mm. Bending at less than the
minimum bending radius may result in damage to the insertion tube.