Chapter 3 Preparation and inspection before observation
3 Check that there is no dirt or water on the objective lens of the
insertion tube's distal end. If dust or water is present, wipe them
clean with a soft piece of gauze, a cotton swab, or sweep them off
with a brush, grasping the hard part of the distal end. The best
results may be obtained using a commercially available absolute
ethanol or isopropyl alcohol.
CAUTION
Using the objective lens of the insertion tube's distal end while dirt or water is present may
cause damage to the optical adapter or the insertion tube's distal end.
4 Check that the distal end of the insertion tube is not deformed,
loose, or that there is not unusual wear on the connecting screw.
CAUTION
Never use the distal end of the insertion tube if any part is loose. Otherwise, the loose part
may fall into the inspection object.
The angulation section is composed of precision parts. Do not pull the distal end of the
insertion tube, do not grasp/bend with a strong force. Otherwise, the parts in the angulation
section may be damaged.