OMNI 4000/7000 Installation Guide Rev H
Page A-6
Jumper Settings
A.3.3 B Module Process Input Type Pulse
The B module (or E/D module) is required when connecting to digital densitometers. With the B
module, Analog Output #2 is never available, because the periodic time function uses the internal
timer counter that is normally used to generate the second analog output.
To configure a B module for Pulse Inputs, set the jumpers according to these instructions.
Configure Inputs 1 and 2 for one of three analog signal types:
For 1-5 VDC Input, remove both RTD and 4-20 jumpers.
For 4-20 mA Input, install the 4-20 jumper.
For RTD Input, install the RTD jumper.
Configure Input 3 for pulse signal types:
Remove all four “A” jumpers and install the two “P” jumpers.
Remove the 4-20 jumper.
Configure Input 4 for pulse signal types:
Remove all four “A” jumpers and install the two “P” jumpers.
Signal coupling is instrument dependent.
For AC coupled, install the two AC jumpers.
For DC coupled, install the DC jumper.
Signal threshold is instrument dependent:
For a 1.5v threshold, remove the HI jumper.
For a 3.5v threshold, install the HI jumper.
Remove the 4-20 jumper.
Refer to Figure A-4 for default input and output jumper settings: Inputs Ch1 = 4-20mA,
Ch2 = 4-20mA, Ch3 = 4-20mA, Ch4 = AC coupled Pulse and Output Ch2 = RTD2.