Operator’s Manual – OPTI CCA-TS 6-5
6 MAINTENANCE
Fig. 6-10 HbCal Report
OPTI Medical OPTI CCA-TS
HbCal Report
DD-MMM-YY HH:MM
S/N: XXXX
Version: ABCX.XX
Operator ID:
HbCal Type: tHb-Calibrator
HbCal LOT: XXXXXX
HbCal Date: DD-MMM-YY
Calibration Results:
Meas’d Cal’d
tHb 12.9 13.0
S02(%) 74.6 74.9
Calibration Factors:
OLD NEW
F1 1.023 1.014
F2 1.087 1.080
F3 1.089 1.094
F4 0.000 0.000
F5 0.000 0.000
The unit will now begin printing the tHb Calibration
Report showing both the old and new calibration
results and calibration factors (Fig. 6-10).
13. After completion of Hb Calibration, it is
necessary to delete and reload the current SRC
data for each level. (Refer to section 3.3.1.1
Setting up the Standard Reference Cassette).
14. Press (Fig. 6-11) to proceed to the
<SRC Setup> menu.
15. Verify performance by now running SRC
measurements and 3 levels of OPTI Check.
Fig. 6-11 Delete and rescan SRC data
Delete old SRC Data and
rescan SRC Barcodes
OK
Delete old SRC Data and
rescan SRC Barcodes
OK