EasyManua.ls Logo

OPTIMedical OPTI CCA-TS - Page 109

OPTIMedical OPTI CCA-TS
271 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6-4 Operator’s Manual – OPTI CCA-TS
6 MAINTENANCE
Fig. 6-9 tHb Calibration
9. Gently wipe both sides of the tHb-Calibrator
Cassette with a clean dry cloth and examine it to
ensure it is clean. Insert it into the chamber and
press down to properly seat the cassette
(Fig. 6-7).
10. Close the sample chamber cover (Fig. 6-8).
After the cover has been closed, the instrument will
automatically detect the presence of the calibrator
cassette and begin calibration (Fig. 6-9).
11. After the calibration is complete you will be
prompted to open the sample chamber cover and
remove the cassette.
12. Place the calibrator cassette back into its pouch
immediately after removal from the instrument.
NOTE: Makesuretokeepthecalibratorcassette
withtheinstrumentatalltimes.
Fig. 6-8 Close Cover
Fig. 6-7 Wipe and Insert Cassette

Table of Contents