Operator’s Manual – OPTI CCA-TS 6-3
6 MAINTENANCE
5. Enter the lot number of the tHb-Calibrator
cassette located on the top surface of the cassette
and press (Fig. 6-4).
6. At the prompt open the SMC cover by pressing
the button (Fig. 6-5).
7. Gently clean the optics window and the inside
top cover of the sample chamber with a soft lint
free cloth (Fig. 6-6). Press
.
Fig. 6-5 Open Cover
Fig. 6-4 Enter Lot Number
Fig. 6-6 Clean Optics and
Inside of Cover
Clean Optics and
Inside of Cover.
Optics
OK
Clean Optics and
Inside of Cover.
Optics
OK