Ca-2 Operator’s Manual – OPTI CCA-TS
Interferences
The OPTI CCA-TS Ca
++
sensor does exhibit an interference from bisulfate and phenylacetic acid.
Heparin salts are the only acceptable anticoagulants. Other anticoagulants such as citrate, EDTA, oxalate,
and uoride cause signicant interferences to the electrolyte sensors.
The following exogenous interferents were quantied in tonometered plasma, showing interferences to
dyes which typically have short half-lives within the body before being metabolized by the liver.
Substance amount Ca
++
change mmol/L
Sodium uorescein 26 mg/dL unstable
Cardio (indocyanine) green 0.5 mg/dL +0.01
Methylene blue 25 mg/dL unstable
Only clear, uncolored quality control materials, such as OPTI CHECK or OPTI CHECK PLUS brand
aqueous controls should be used with the OPTI CCA-TS system. Colored materials, including prociency
testing materials, may interfere with the ion measurement, or fail to be properly aspirated.
Reproducibility
Typical Within-Run (Swr), Between-Day (Sdd) and Total (ST) Precision is determined from 1 run per
day with 2 replicates per run for 20 days on each of two OPTI CCA instruments. Ionized calcium values
are expressed in mmol/L.
Material mean Swr (CV%) Sdd (CV%) ST (CV%)
OPTI CHECK PLUS
(aqueous control solution)
Level 1 1.26 0.001 (0.1) 0.007 (0.6) 0.007 (0.6)
Level 2 1.16 0.002 (0.2) 0.010 (0.9) 0.011 (0.9)
Level 3 0.79 0.001 (0.1) 0.005 (0.7) 0.005 (0.7)
Serum 0.64 0.001 (0.1) 0.01 (1.8) 0.01 (1.8)
Reduced Bovine Hemoglobin 0.81 0.001 (0.2) 0.007 (0.9) 0.007 (0.9)
Solution
All specic performance characteristics tests were run with default instrument calibration and after
normal recommended equipment quality control checks were performed (see Operator’s Manual).
Specimens at each level were analyzed in replicates of two for 20 days. The within-run and
between-day standard deviations were calculated by the analysis of variance method.