Chapter 3 - How to...
3.4.8 Troubleshooting
If you have any issues using your device please refer to the following table and instructions.
Additional information and resources are available on the Optos website. Go to optos.com and register to
access the information and resources.
Detailed instructions on using OptosAdvance can be found in the help files for each application.
Note
If you require further assistance, please contact your Optos representative, see Contact us on page71.
When contacting Optos you will need to supply your site number, see What information do I need when
contacting Optos? on the facing page.
TABLE 6: List of problems and solutions
Problem Possible cause Solution
Capture mode is
disabled.
Failure of the system self-test
disables capture.
Open the diagnostics page and run the self-test
feature.
Cannot access the
user login list on the
scan head.
Password error, image server
is NOT switched on or scan
head is NOT configured with
the correct image server IP
address.
Check your user name and password, see Why can't
Ilogin? on the facing page.
Check the image server is switched on and running,
see How do I restart the image server? on page62.
Check the image server network address and ensure
the scan head is on the same network and configured
to connect to the correct address.
Patient position
feedback NOT
showing on some
eyes.
Ambient lighting too bright;
irregular pupil shape;
unmanaged ptosis.
Ensure the patient’s pupil is centered on the cross
hairs on the touch screen.
Cannot align the
patient with the hand
control.
Equipment failure; device set
up incorrect.
Ensure the patient’s pupil is centered on the cross
hairs on the touch screen. Contact Optos for
assistance.
Imaging when the
patient position
feedback is green
does NOT produce
good results.
System virtual point is offset
leading to incorrect feedback.
Compensate by imaging with the alignment target on
red or blue, as appropriate.
Images taken
include large lid and
lash obscuration.
Poor patient position;
unmanaged ptosis.
Review the positioning guidance in this document,
see Capture images on page34. Contact Optos for
additional training if required.
optomap images
show horizontal line
artefacts.
System slit mirror requires
cleaning.
Contact Optos to arrange for a Field Service Engineer
to clean the slit mirror.
optomap images
show small random
artefacts in fixed
position.
Dust on the main mirror, or
main mirror dirty.
Use the recommended cleaning materials and
procedure for cleaning the main mirror, see Cleaning
the main mirror on page58.
If the issue persists contact your Optos
representative.
Page 60 of 75 Part Number: G108707/8GME
English Copyright 2018, Optos plc. All rights reserved.