Chapter 4 - Technical Specification
4.2.2 Internal Laser warning information
Location of Laser warning Warning information
Positioned in the inside rear of
the device.
CAUTION - CLASS 3B LASER RADIATION
WHEN OPEN AVOID DIRECT EYE EXPOSURE
Positioned on the beam cover.
VISIBLE AND INVISIBLE LASER RADIATION
AVOID DIRECT EXPOSURE TO BEAM
CLASS 3B LASER PRODUCT
Positioned on the SLD cover
WARNING - LASER RADIATION
AVOID EXPOSURE TO BEAM
CLASS 3B LASER PRODUCT
4.3 Electromagnetic compatibility (EMC)
The P200TE is intended for use in the electromagnetic environment as specified in the following tables. The
customer or the user of the device should assure that it is used in such an environment.
4.3.1 Guidance and manufacturer declaration - electromagnetic
emissions
TABLE 13: List of electromagnetic emission compliance
Emissions test
IEC60601
compliance
test level
Electromagnetic environment-guidance
RF emissions
CISPR 11
Group 1
The P200TE uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are NOT likely to
cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The P200TE is suitable for use in all establishments, including
domestic establishments and those directly connected to the public
low-voltage power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC61000-3-3
Pst
Dmax
4.3.2 Guidance and manufacturer declaration electromagnetic
immunity
TABLE 14: List of electromagnetic immunity compliance
Immunity test
IEC60601 compliance test
level
Electromagnetic environment-guidance
Electrostatic
discharge
IEC61000-4-2
±6 kV contact
±8 kV air
Floors should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative humidity
should be at least 30%.