Chapter 4 - Technical Specification
Where wheelchair accessibility is required the table must meet wheelchair accessibility standards.
4.1.7 Battery
CAUTION
Do NOT remove the scan head covers. Only Optos trained technical personnel are
permitted to service the device.
The device contains a 3V Lithium coin battery.
4.1.8 Environmental considerations (ambient)
Do NOT operate the device in an environment that exceeds the environmental limits, see Warnings and
cautions on page11.
TABLE 12: Environmental ranges
Status Temperature
Relative Humidity
(Non-condensing)
Atmospheric Pressure
Normal use +10°C to +35°C 75% 70kPa to 106kPa
Storage -10°C to +50°C 95% 70kPa to 106kPa
Transport -10°C to +60°C 95% 50kPa to 106kPa
Note
Device is DEHP free.
4.2 Laser safety designation (IEC 60825-1)
CAUTION
The P200TE is a Class 1 laser device at the eye, and complies with IEC/EN 60825-
1:2014 and 21 CFR1040.10 and 1040.11 except for deviation pursuant to Laser Notice 50,
dated June 24, 2007.
Based on current scientific knowledge, a Class 1 laser device can be considered as safe
by engineering design, and safe under reasonably foreseeable conditions of operation.
Wavelengths:
SLO: 532nm, 635nm
SLD Center wavelength: 840nm ±10 nm
Resolution:
Typical Axial Resolution: <10 micron, Digital on-screen <6 micron
Transverse Resolution: 20 micron (in tissue)
optomap plus: 14 micron
4.2.1 External Laser warning information
Location of Laser warning Warning information
Positioned on the rear scan
head cover.
CLASS 1 LASER DEVICE
Page 66 of 75 Part Number: G108707/8GME
English Copyright 2018, Optos plc. All rights reserved.