EasyManua.ls Logo

Optos P200TE - 4.3.3 Recommended separation distances

Optos P200TE
374 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 4 - Technical Specification
Immunity test
IEC60601 compliance test
level
Electromagnetic environment-guidance
Radiated RF
IEC61000-4-3
3V/m
80MHz to 2.5GHz
d=1.2√P 80MHz to 800MHz
d=2.3√P 800MHz to 2.5GHz
Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic survey,
a
should be
less than the compliance level in each frequency range.
b
Interference may occur in the vicinity of equipment
marked with:
Note 1: At 80MHz and 800MHz the higher frequency range applies.
Note 2: These guidelines may NOT apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment in the location due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which the
device is used exceeds the applicable RF compliance level above, the device should be observed to verify
normal operation. If abnormal performance is observed, additional measures may be necessary, such as
reorienting or relocating the device.
b Over the frequency range 150kHz to 80MHz, field strengths should be less than 3V/m.
4.3.3 Recommended separation distances
This section lists the recommended separation distances between portable and mobile communication
equipment and the device.
The P200TE is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled.
The customer or the user of the P200TE can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communication equipment (transmitters) and the P200TE
as recommended below, according to the maximum output power of the communications equipment.
TABLE 15: List of recommended separation distances
Rated maximum output power
of transmitter
(W)
Separation distance according to frequency of transmitter
150kHz to 80MHz
d=1.2√P
80MHz to 800MHz
d=1.2√P
800MHz to 2.5GHz
d=2.3√P
00.1 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1.0 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
Part Number: G108707/8GME Page 69 of 75