Communications Interfaces
6400
12
71
Test Data Files
6400 Data File Naming Convention
Test files are stored in the calorimeter with a unique
file name. See the following table for details.
Test Type Filename
Preliminary Standardization <ID>.std.plim.csv
Final Standardization <ID>.std.finl.csv
Preliminary Determination <ID>.det.plim.csv
Final Determination <ID>.det.finl.csv
Pre-weigh <ID>.---.pwgh.csv
6400 Calorimeter Run Data Template
Each test file contains information separated by
commas. See the following table for a list of the
data contained within the file. In the data directory
on the calorimeter there is also a RunDataTem-
plate.csv file that contains column headings for the
test file data and a RunDataTemplate.txt file that is
a duplicate of the table below.
Field Description
SampleID char[16]
Timestamp MM/DD/YY HH:mm:ss
Mode 0 = undetermined (preweigh),
1 = determination,
2 = standardization
Method 0 = undetermined (preweigh),
1 = equilibrium,
2 = dynamic
State 0 = preweigh,
1 = preliminary,
2 = final
Units 0 = MJ/kg,
1 = Btu/lb,
2 = cal/g,
3 = J/kg,
4 = other
UnitMultIfOther unit mult in effect at time of
report, if Units = 4 (other)
BombID [1,4]
BombEE bomb energy equivalent
Field Description
SampleWt sample weight
SpikeWt spike weight
Fuse fuse value
FuseFinal fuse value is final (0 if not final, 1
if final)
Acid acid value
AcidFinal acid value is final
Sulfur sulfur value
SulfurFinal sulfur value is final
Hydrogen hydrogen value (net calc option)
HydrogenFinal hydrogen value is final (net calc
option)
MAD moisture as determined value
(dry calc option)
MADFinal moisture as determined value is
final (dry calc option)
JacketTemp jacket temperature
InitTemp initial temperature
DeltaT temperature rise
HOC gross heat of combustion
NetHOC net HOC (if net calc option en-
abled)
DryHOC dry gross HOC (if dry calc option
enabled)
DryNetHOC dry net HOC (if both dry and net
calc options enabled)
Oxygen oxygen value (net calc option)
OxygenFinal oxygen value is final (net calc
option)
Nitrogen nitrogen value (net calc option)
NitrogenFinal nitrogen value is final (net calc
option)
MAR moisture as received value (dry
calc option)
MARFinal moisture as received value final
(dry calc option)
DryNetHOC_AR net HOC as received (if all dry and
net values provided)
BombName user friendly name for bomb

Table of Contents