Calibration
Lasair II Particle Counter Operations Manual Page 10-5
Calibration
Calibration should be performed once each year. The Lasair II particle counter
performs a calibration date check each time it is energized. If the current date is
within 30 days or less of the anniversary of the last calibration, a warning message
will be displayed.
The warning is displayed only on the main LCD display. It is not shown on the Web
server. If the calibration date past due, (365 days since the last calibration date), the
message will include the number of days the calibration is past due. The instrument
also notifies the user if no calibration date is saved.
Other Maintenance or Operational Problems
For other maintenance, call Particle Measuring Systems first for advice and
technical information at 1-800-557-6363. There is nothing within the Lasair II
particle counter case that requires user maintenance.
If you are experiencing operational problems that you are unable to solve, see
Appendix A, "Troubleshooting".
Figure 10-1: Calibration Warning Dialog Box

Table of Contents

Other manuals for Particle Measuring Systems LASAIR II 550L

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Particle Measuring Systems LASAIR II 550L and is the answer not in the manual?

Particle Measuring Systems LASAIR II 550L Specifications

General IconGeneral
BrandParticle Measuring Systems
ModelLASAIR II 550L
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals