Setting Analog Alarms
Lasair II Particle Counter Operations Manual Page 5-17
You can set a Minimum value, a Maximum value, or both for each analog variable
in the list. You must also enable those alarms by entering a bullet in the 1/0 field
next to the Min/Max field.
To set minimum and/or maximum values for an analog probe:
1 Open the Analog Alarm Setup Screen.
2 Press Enter to activate Edit Mode.
3 Tab to the first Min value field in the list. Use the directional arrow buttons on
the control panel to move around the list of variables.
4 Press Enter to initiate the ability to change an alarm limit in a Min or Max field.
5 Use the editing icon softkeys and the up and down arrows to set or change the
alarm limit.
6 Press Enter to record that value and to move to the 1/0 radio button.
7 Press Enter again to insert a bullet in the 1/0 radio button. This alarm is now
enabled at the alarm limit you have selected.
Figure 5-9: Alarm Setup Screen, Analog Alarms Page

Table of Contents

Other manuals for Particle Measuring Systems LASAIR II 550L

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Particle Measuring Systems LASAIR II 550L and is the answer not in the manual?

Particle Measuring Systems LASAIR II 550L Specifications

General IconGeneral
BrandParticle Measuring Systems
ModelLASAIR II 550L
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals