5
HD11 XE Getting Started
4535 612 62651
119
Cleaning and Maintaining the System
Service Documentation
Because the system is a piece of medical equipment that contains several circuit
boards, extensive service diagnostics, and complex operating software, Philips rec-
ommends that only trained personnel service the system.
Service manuals are not shipped with each system. However, you can purchase a
service manual from your Philips representative. It contains parts lists, a brief the-
ory of operation, error codes, and preventive maintenance and safety testing
instructions.
Monitor As necessary Clean as instructed. “Cleaning and
Disinfecting System
Surfaces” on
page 115
Air filter 3 months or
more often
Clean as instructed. “Cleaning the Air
Filter” on page 121
Trackball 3 months Clean as instructed. “Cleaning the
Trackball” on
page 122
VCR 6 months Clean video heads and tape path. “VCR” on page 123
Video
printers
6 months Clean print head, rollers, and
sensors.
“Video Printers” on
page 123
Fan 1 year Check for good airflow without
excessive noise.
“Service
Documentation” on
page 119
Safety 1 year Perform ground impedance/leakage
tests.
Table 5-1 Recommended Frequency of Maintenance Tasks (Continued)
Item Frequency Task See