Philips HD11 XE - Figure 12-3 Test 1-Transducer Leakage Current Test (Source)

Philips HD11 XE
278 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Intraoperative Transducers
HD11 XE Getting Started
4535 612 62651
12
254
Figure 12-3 Test 1—Transducer Leakage Current Test (Source)
Key:
C Stray capacitance from the ultrasound system power wiring to the system
grounded metal chassis (1 to 3 M reactance)
Z Impedance between the metal parts of the transducer and a test electrode
placed in the bucket of saline solution (about 850 k with an intact outer
insulating layer, 500 with a hole in the layer)
A Microammeter to measure leakage current from the transducer to the Earth
Lift Ground through Z to the test electrode
e Line power source, either 110 Vac or 220 Vac
i Current caused by e and stray capacitance, and optionally Z
S1 Open Earth Lift Ground switch
S2 Line polarity switch
i
Ultrasound system
(metal chassis)
Saline
A
L
S2
S1
N
E
C
Test container
(nonconductive)
Test electrode
Intraoperative transducer
immersed in saline solution
Z
e
i