Transducer Care and Maintenance
To install the ergonomic grip
1. Insert the cable through the slot in the grip.
2. Align the slot with the orientation indicator.
3. Slip the transducer head into the grip.
CAUTION
During installation, avoid transducer lens damage from fingertips and nails by
pressing the transducer into the grip with the palm of your hand.
To clean the ergonomic grip
1. Remove the grip and wipe off excess gel.
2. Wash with soap and running water, rinse thoroughly, and air dry.
3. Alternatively, it can be low level disinfected by wiping or spraying the grip with
a bleach (10%), a quaternary ammonium, or a 70% isopropyl alcohol (rubbing
alcohol) product for the recommended contact time. Do not allow the disinfec-
tant to dry on the grip. Rinse and air dry.
Storing Transducers
This section provides information on storing transducers for transport, and daily
and long-term storage.
Storage for Transport
If a carrying case is provided with your transducer, always use the carrying case to
transport the transducer from one site to another.
Follow these guidelines to properly store transducers for transport:
Make sure that the transducer is clean and disinfected before placing it in the
case to avoid contaminating the foam that lines the carrying case.
Place the transducer in the case carefully to prevent kinking of the cable.
•Before closing the lid, make sure no part of the transducer is protruding from
the case.