Table 6-11 ECG/Physio Input Parameters and Values
Element Parameter Value
ECG input 3-dB bandwidth 1.0 Hz ± 0.5 Hz to 30 Hz
± 6Hz
Minimum input signal to
trigger R-wave detector
0.25 mV
Common mode rejection > 90 dB at 60 Hz
Forward leakage to ground < 6.5 µΑ at 260 VRMS
Forward leakage to mains < 40 µΑ at 264 VRMS
Reverse leakage to ground < 6.5 µΑ at 240 VRMS
Reverse leakage to mains < 40 µΑ at 264 VRMS
Spark gap breakdown voltage 500 to 680 VRMS
Input for full scale at 100%
gain setting in Dual M-Mode
±5.5 mV P-P
Aux ECG input Input for full scale at 100%
gain setting
1.25 V P-P ± 0.25 V
3-dB bandwidth DC to 100 Hz min
Maximum input signal ± 4 V
Aux physio
input
Input for full scale at 100%
gain setting in Dual MMode
3.0 V P-P
3-dB bandwidth DC to 100 Hz min
Maximum input signal 4V P-P