2. Press Door open / assistant call.
A pop-up window opens.
3. Adjust the door open / assistant call time.
The minimum value is 0 seconds, the maximum 250 seconds. The value
is adjusted in steps of 1 in value range 0 - 30, and in steps of 5 in value
range 30 - 250. The default value is 5.
4. Press OK.
The pop-up window closes.
25.9 Clock
About this task
NOTE
If Planmeca Romexis Clinic Management software is used together with this
dental unit, you do not need to adjust the clock settings. The dental unit
clock is synchronised with the Planmeca Romexis clock every time the
dental unit is booted up.
Steps
1. Press Program.
2. Press Clock.
A pop-up window opens.
3. Select the clock programming view by pressing Clock at the bottom of
the programming window.
4. Press 12/24 to change the mode (12-hour / 24-hour clock).
5. Adjust the time.
6. Press OK.
25 Programming
198 Planmeca Compact i User's manual