25.11 Adjusting maintenance functions
About this task
You can adjust the following maintenance functions in the
Program -
Maintenance
window:
disinfectant concentration
duration of short flushing
duration of long flushing
duration of WCS final flushing.
Alternatively, the values can be adjusted in the service modes, contact your
Planmeca dealer.
Steps
1. Press Program.
2. Press Maintenance.
The
Program - Maintenance
window opens.
3. Press Edit next to any item to open a new window where you can adjust
the disinfectant concentration and/or duration of maintenance programs.
A pop-up window opens.
4. Adjust the value by dragging the slider handle.
Dragging the slider handle to the right increases the value and dragging
it to the left decreases the value.
Default value Value range
Disinfectant concentration 2 % 1 - 3 %
Duration of short flushing 30 s / hose 10 - 30 s / hose
Duration of long flushing 2 min total 2 - 9 min total
Duration of WCS final
flushing
4 min total 4 - 15 min total
5. Press OK.
The pop-up window closes.
25.12 Confirming suction disinfectant container fill
About this task
When you have filled the suction disinfection container, you must confirm the
filling.
25 Programming
200 Planmeca Compact i User's manual