2. Press Program.
3. Press Chip blow.
A pop-up window opens.
4. Adjust the flow rates for water and air.
The minimum value of both parameters is 0 (no flow) and maximum 100.
The adjustment step is 1 in the value range 0 - 35% and 5 in the value
range 35 - 100%.
5. Press OK.
The pop-up window closes and the new values are displayed on the
control panel.
25.4.5 Instrument light
Steps
1. Activate the instrument.
2. Press Program.
3. Press Instrument light.
A pop-up window opens.
4. Adjust the light intensity.
The minimum value is 70% of the maximum intensity and the maximum
100%, and the adjustment step is 2.
After reaching the minimum value the light is switched off.
5. Press OK.
The pop-up window closes and the new values are displayed on the
control panel.
25.4.6 Bien-Air MCX micromotor
About this task
Changing the instrument spray, chip blow and LED values for the Bien-Air
MCX micromotor follows the normal programming pattern. For more
25 Programming
User's manual Planmeca Compact i 183