Red indicator light 100 % full
In addition, help message H 1 is displayed when the amalgam collector is 95
% full, and H 2 when the amalgam collector is 100 % full.
DÜRR CS1 suction system
Check the deposit cup monthly and replace it when it is 70 % full.
Wet suction system
Check the deposit cup monthly and replace it when it is 70 % full.
29.5.7 Clean-water bottle
Use a bottle brush and a mild soap- and water solution to clean the clean-
water bottle once a week.
29 Cleaning and disinfection