6. Confirm the prefilter change by pressing Reset on help message H90.2.
Alternatively, the change can be reset in the
ActiveAqua maintenance
window, see section "Resetting prefilter counter" on page 256.
A confirmation message is displayed.
7. Confirm the reset by pressing OK.
8. Wait for 90 s while ActiveAqua circulates water to bleed the prefilter.
30.4 Resetting prefilter counter
About this task
The prefilter's lifetime depends on the water hardness setting and the clean
water consumption, the maximum change period for the prefilter being 25
weeks.
When the prefilter has been changed, you must confirm the change either by
pressing the prefilter reset button on help message H90.2 or from the
ActiveAqua maintenance
window (described below).
Steps
1. Press Program.
2. Press About this unit.
30 Planmeca ActiveAqua
256 Planmeca Compact i User's manual