NOTE
When the dental unit is equipped with hanging-tube instruments, dry the
instrument holders properly after cleaning. Wet surfaces might disturb
instrument recognition.
NOTE
The instrument console and control panel must be completely dry before
covering them with a protective cover.
29.3 Instruments
Clean and service the instruments according to the information supplied with
the instrument.
NOTE
After cleaning the instrument, let the oil run from it for at least 10 minutes
before replacing it on the console.
Do not use lubricating oil on the Morita TORX micromotor (with the exception
of its O-rings).
Throw away the intraoral camera’s disposable hygiene sleeve after use.
29.4 Instrument console
29.4.1 Oil collector
Instrument console with balanced instrument arms
The oil collector underneath the instrument console has to be emptied and
cleaned monthly. Remove the oil collector by turning it counter-clockwise as
shown in the picture below.
Instrument console with hanging-tube instruments
The oil collector underneath the instrument console has to be checked
monthly and emptied when full.
29 Cleaning and disinfection
238 Planmeca Compact i User's manual

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Planmeca Compact i5 and is the answer not in the manual?

Planmeca Compact i5 Specifications

General IconGeneral
BrandPlanmeca
ModelCompact i5
CategoryDental equipment
LanguageEnglish

Related product manuals