Separator system related error messages
ERROR
CODE
EXPLANATION
E63.1 Water management system (WMS) seems to be installed,
although it should not be (according to service mode).
E63.2 No Water management system (WMS) is installed, although it
should be according to service mode.
32.3.10 Patient chair related error messages
Patient chair related error messages
ERROR
CODE
EXPLANATION
E64 Lift motor position potentiometer or its cable is faulty or
disconnected.
E65 Lift motor does not run or no signal from position
potentiometer.
E65.1 Lift motor does not run. EmeStop relay PCB/cables or position
sensor/cable may be faulty.
E66 Lift motor or its position potentiometer cable has a wrong
connection.
E67 Backrest position potentiometer or its cable is faulty or
disconnected.
E68 Backrest motor does not run or no signal from position
potentiometer.
E68.1 Backrest motor does not run. EmeStop relay PCB/ cables or
position sensor/ cable may be faulty.
E69 Backrest motor or its position potentiometer cable has a
wrong connection.
E70.1 Lift motor error.
E70.2 Backrest motor error.
32.3.11 Main CPU related error messages
Main CPU related error messages
ERROR
CODE
EXPLANATION
E72.1 Program (Flash-EPROM) error. Main software is corrupted.
Replace Flash- EPROM.
E72.2 Program (Flash-EPROM) error. Boot software is corrupted.
Replace Flash- EPROM.
E72.3 Program (Flash-EPROM) error.
Update MCB software.
E72.4 Data transfer error while updating main PCB SW from control
panel memory device.