EasyManua.ls Logo

Planmeca Compact i5 - Automatic legrest

Planmeca Compact i5
326 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10.6.5 Adjusting sterile water volume
Use the clamp on the sterile water tube to adjust the water volume. Push the
slide forward to reduce the volume, and backward to increase the volume.
10.6.6 Removing sterile water bag holder (OP deliveries)
The sterile water bag holder that is mounted on the column can be removed
from its mount as follows:
1. Loosen the screw on the mount with a 4 mm Allen key.
2. Remove the sterile water bag holder.
3. Tighten the screw.
10 Instrument system
56 Planmeca Compact i User's manual

Table of Contents

Related product manuals