Calibration 
 
CIRAS-2 Operator's Manual Version 2.04 
- 144 - 
through a 30 cm tube.  The flow rate from the gas cylinder should be set so that there is a small excess, 
say 100-200 ml/minute, flowing to atmosphere.  The total flow is then 300 to 400 ml/minute. 
 
ON  NO  ACCOUNT  PLACE  THE  VENT  TUBE  INTO  WATER  TO  INDICATE  SURPLUS  FLOW.    IF 
CIRAS-2  INGESTS  WATER,  IT  IS  LIKELY  THAT  THERE  WILL  BE  DAMAGE  TO  THE  CELL 
SURFACES,  THE  DETECTOR  WINDOWS,  THE  IR  SOURCE,  THE  PUMPS  AND  THE  MASS 
FLOWMETERS. 
 
Alternatively:  
 
Fill a gas sample bag with the calibration gas direct from the cylinder.  Now attach the gas sample bag 
inlet to the REF AIR IN and the CIRAS-2 will draw out a sample without any risk of pressurization. 
Please note: Be sure to initially partially fill and empty the bag to ensure that the sample is truly the 
calibration gas. 
 
If possible, the standards should have a concentration slightly greater than the usual measurement.  It is 
better not to calibrate with a concentration much smaller than the measurements (i.e. do not use a 300 
ppm standard when measuring at 2,000 ppm. 
 
H
2
O Calibration sources and connections 
 
For  H
2
O  calibration, an  air  supply of  known water  vapor  concentration is required (mb). There  are  a 
number of ways that this can be obtained that are listed in the technical manual. 
 
It is however assumed here that you will be using the water vapor calibrator supplied with your system.  
Instructions for use of the H
2
O calibrator can be found in the CIRAS-2 Technical Manual. 
 
PP Systems‟ Water Vapor Calibrator 
The calibrator consists of an insulated container filled with ferrous sulphate salt (FeSO
4
.7H
2
O). The unit is 
fitted with thermistor temperature sensors measuring ambient air temperature and the temperature of the 
air emerging from the FeSO
4
.7H
2
O.  The water vapour pressure above the FeSO
4
.7H
2
O is a function of 
the salt temperature as shown in the table in the CIRAS-2 Technical Manual. The salt can humidify many 
litres of air, and it should last many years as it is being used in a re-circulating mode.  The degree of 
exhaustion of the salt can be determined by its weight as shown on the test certificate beneath its cap or 
on the test sheet supplied with your documentation. For an in-depth discussion please see: 
 
Ref  Parkinson K.J. and Day W. (1981) Water Vapour Calibration using Salt Hydrate Transitions. J. Expt 
Bot. 32, 411-418. 
 
  The calibrator should be both stored and operated in a relatively stable temperature environment. 
  The  calibrator  and  CIRAS-2  should  be  placed  together  in  a  stable  temperature  to  allow  their 
temperatures  to  equilibrate  for  several  hours  before  use.    If  the  equilibrator  is  warmer  than 
CIRAS-2 then there could be a risk of condensation. 
  The electrical connection is to the PLC connector. 
  The REF air inlet should be connected to the bottom pipes on the calibrator and the REF and AN 
outlets to the two top pipes. 
  Pipe  connections  should  not  be  made  until  it  has  been  ascertained  that  there  are  no  error 
messages from CIRAS-2 that indicate the calibrator temperature is too high.