Recalculate Data
CIRAS-2 Operator's Manual Version 2.04
- 81 -
Recalculate Data
There may be situations when you need to recalculate your data, such as post-determination of actual leaf
area, incorrect entry of PAR value, etc. To recalculate data click on File - Open and select the data file
that you want to recalculate and click Open.
The Save Recalculated Data to… dialog will appear. The selected file will be automatically given a .mod
extension. The original file (.dat extension) will remain unaltered. Click Save to continue. The CIRAS-2
Photosynthesis Recalculation dialog will appear. Initially, data from the first record stored is displayed.
To recalculate the data for the displayed record, check off the parameter(s) needing to be changed under
Recalculation Values and enter the desired value(s). Alternatively, click on Skip to move to the record
requiring recalculation and make the desired changes. After completing changes for a specific record,
click OK to accept the changes and to have the data recalculated.