If
[Sensor not dry]
continues to be displayed
despite the conductivity sensor having been
dried, then you will have to wait some time until
the controller has detected the sensor as dry.
Once you have selected the sensor type, the
prompt automatically appears asking whether
the sensor parameters (zero point) have to be
determined. You can initiate this prompt man‐
ually as follows:
Continuous display ➨ Menu ➨ or
[Measurement]
or
[Measuring channel X Conductivity]
or
[Sensor parameter adjustment]
.
1. Use the arrow keys to select
[Automatically determine sensor param‐
eters]
.
2. Continue with .
ð
You will see the display showing
[Sensor dry]
and
[Automatically determine sensor
parameters]
.
3. Continue with .
ð
You will see the display with the
message
[Sensor parameters are automati‐
cally determined]
.
The sensor parameters are auto‐
matically carried over.
Calibration
116