Selecting the calibration process
1. Continuous display ➨
ð
The Calibration menu is displayed,
you may need to select
[Channel 1]
or
[Channel 2]
, depending on the
measuring channel on which the
pH measurement is performed.
2. Press
CAL pH
Buffer manufacturer
Calibration process
A1025
Off
Buffer detection
2 point
requirement
Buffer value 1
Buffer value 2
Buffer temperature
ProMinent
pH 7
pH 4
Fig. 37: Selecting the calibration process
ð
The menu for selecting the calibra‐
tion process appears.
3. Use the arrow keys to select the
required menu item and press
ð
The input window appears and you
can make the necessary settings
for your process
4. Use the arrow keys to select the calibra‐
tion process and press
5. Continue with
ð
You can now start your chosen cali‐
bration process.
10.1.2 2-Point Calibration of the
pH Sensor (CAL)
Correct sensor operation
Correct measuring and metering is
only possible if the sensor is
working perfectly.
Refer to the sensor's operating
instructions.
2-point calibration is strongly rec‐
ommended and is preferable to
other methods.
The sensor needs to be removed
and refitted in the in-line probe
housing for calibration. Refer to
the operating instructions for your
in-line probe housing.
Calibration
85

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ProMinent DULCOMETER diaLog DACb and is the answer not in the manual?

ProMinent DULCOMETER diaLog DACb Specifications

General IconGeneral
BrandProMinent
ModelDULCOMETER diaLog DACb
CategoryController
LanguageEnglish

Related product manuals