Selecting the calibration process
1. Continuous display ➨
ð
The Calibration menu is displayed,
you may need to select
[Channel 1]
or
[Channel 2]
, depending on the
measuring channel on which the
pH measurement is performed.
2. Press
CAL pH
Buffer manufacturer
Calibration process
A1025
Off
Buffer detection
2 point
requirement
Buffer value 1
Buffer value 2
Buffer temperature
ProMinent
pH 7
pH 4
Fig. 37: Selecting the calibration process
ð
The menu for selecting the calibra‐
tion process appears.
3. Use the arrow keys to select the
required menu item and press
ð
The input window appears and you
can make the necessary settings
for your process
4. Use the arrow keys to select the calibra‐
tion process and press
5. Continue with
ð
You can now start your chosen cali‐
bration process.
10.1.2 2-Point Calibration of the
pH Sensor (CAL)
Correct sensor operation
–
Correct measuring and metering is
only possible if the sensor is
working perfectly.
–
Refer to the sensor's operating
instructions.
–
2-point calibration is strongly rec‐
ommended and is preferable to
other methods.
–
The sensor needs to be removed
and refitted in the in-line probe
housing for calibration. Refer to
the operating instructions for your
in-line probe housing.