EasyManuals Logo

ProMinent DULCOMETER diaLog DACb User Manual

ProMinent DULCOMETER diaLog DACb
192 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #86 background imageLoading...
Page #86 background image
Defining buffer detection
There are 2 buffer detection options
with 2-point calibration.
[Presetting]: select 2 buffers from the 4
possible buffer sets. During calibration,
adhere to the selected order e.g. Buffer
value 1: pH 7 and Buffer value 2: pH 4:
ProMinent
®
(pH 4; 7; 9; 10).
(default setting)
NBS/DIN 19266 (pH 1; 4; 7, ; 9).
DIN 19267 (pH 1;4; 7; 9; 13).
Merck + Riedel
®
(pH 2; 4; 7; 9;
12).
The buffer sets differ in their pH values
and temperature dependencies set in
the controller. The pH values at the dif‐
ferent temperatures are also printed on
the buffer containers.
[Manual]: enter the buffer value with the
associated temperature into the con‐
troller.
The pH values of the buffer solu‐
tion, at temperatures other than
25°C, are printed in a table on the
label of the buffer bottle.
Select the available buffer.
CAL pH
Buffer manufacturer
Calibration process
A1512
Manual
Buffer detection
2 point
Manual
Buffer value 1
Buffer value 2
Buffer temperature
ProMinent
pH 7
pH 4
Buffer temperature 25.0 °C
Fig. 38: Example: Display in [CAL-Setup]]
Used buffer
Dispose of the used buffer solution. For
more information: refer to the material
safety data sheet for the buffer solution.
Valid calibration values
Valid calibration:
Zero point -60mV…+60 mV
Slope 55 mV/pH…62 mV/pH
Two test containers with a buffer solution are
required for calibration. The pH values of the
buffer solutions must be at least 2 pH values
apart. Thoroughly rinse the sensor with water
when changing the buffer solution.
Continuous display ➨ .
CAL pH
Last calibration
A1016
Zero point
Slope
Calibration process
CAL setup
continue with <CAL>
14:26:07
7,00 pH
2 point
56.64mV/pH
06/04/2013
Fig. 39: pH sensor calibration (CAL)
1. Continue with .
2. Rinse the sensor thoroughly with water
and dry with a cloth (pat dry, don't rub).
3. Immerse the sensor in test container 1
which contains the buffer solution (e.g.
pH 7). Gently move the sensor.
4. Continue with .
Calibration
86

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ProMinent DULCOMETER diaLog DACb and is the answer not in the manual?

ProMinent DULCOMETER diaLog DACb Specifications

General IconGeneral
BrandProMinent
ModelDULCOMETER diaLog DACb
CategoryController
LanguageEnglish

Related product manuals